JMB

eISSN 2093-6338

Table. 2.

Table. 2.

Comparison of ABL90 FLEX PLUS with Beckman Coulter AU5800 at the medical decision level by Passing Bablok regression

Slope (95% CI) Intercept (95% CI) Decision level (mg/dL) 95% CI (mg/dL) Expected value (mg/dL)

Low High
1.000 (0.980 to 1.029) 0.000 (-0.036 to 0.024) 0.600 0.576 0.616 0.600
1.600 1.578 1.628 1.600
6.000 5.890 6.155 6.000
Lab Med Online 2020;10:202~206 https://doi.org/10.3343/lmo.2020.10.3.202
© Lab Med Online