Table. 1.

Precision of the Elecsys AMH plus

QC level Mean SD CV
Within-run Between-run Between-day Total Within-run Between-run Between-day Total
Level 1 1.05 0.0141 0.0129 0.0096 0.0214 1.3 1.2 0.9 2.0
Level 2 5.64 0.045 0.072 0.080 0.117 0.8 1.3 1.4 2.1

Abbreviations: QC, quality control; SD, standard deviation; CV, coefficient of variation.

Lab Med Online 2024;14:106~110 https://doi.org/10.47429/lmo.2024.14.2.106